Setup & Op User guide
34
CALIBRATION SETUP 3
Calibration #: (LB) C - U - L
The active calibration of four possible calibrations (16
calibrations possible with POS upgrade). The elds shown
on the Calibration Setup 1 screen represent the values of the
Calibration # shown. Any edits made to the elds are
attributed to the Calibration # shown.
Linearization Mode: (LB) C
Disables and enables multi-point calibration.
Setup disables multi-point calibration during normal deliveries.
Used when proving and setting multi-point calibrations.
Applied enables multi-point calibration.
Linearization Prover Qty: (FE) C
The metered volume of the prover after a calibration delivery. When the prover volume is entered, the LCR 600 will
prompt you to select a linearization point number and a ow rate. After these values are entered, the LCR 600 will
assign the error accordingly. See calibration instructions.
BEGIN CALIBRATION DELIVERY
Starts a calibration delivery.
Point #: (FE) C
The elds under Point #: are used to calibrate the meter/register at up to 10 different ow rates for each product
type/calibration.
Flow Rate: (FE) C
The ow rate at which the calibration point was measured.
Percent Error: (FE) C
This eld contains the percent difference between the register volume and the prover volume.
Percent Error = [(Prover Qty. - Meter Qty.) ÷ Prover Qty.] x 100
CALIBRATION SETUP
L - Locked
U - Unlocked
C - Weights & Measures Calibration
P - Stop/Pause
F - Factory
Security Clearance
FE - Field Edit Box
LB - List Box
Field Type