4 CALIBRATING THE DI-100 
Please ensure the UNIT and LC (load capacity) commands are entered accordingly, prior to 
performing the calibration. 
4.1 Calibration Mode 
The DI-100 supports two calibration modes: The mV/V calibration or the 2-Point Linear Calibration. 
You may perform the calibration for either mode independently and select the actual calibration 
mode in use. 
4.2  Millivolt-per-Volt (mV/V) Calibration 
Obtain the Rated Output (in millivolt-per-volt, mV/V) from your load cell’s calibration certificate or 
from the label attached on the load cell. This number is the best fit mV/V value determined during 
the factory calibration procedure. 
•  To select the mV/V mode, type: CAL m <Enter> 
•  To enter mVolt calibration, type: MVOLT x.yyyy <Enter> 
Where:  x.yyyy is the mV/V rated output. The parameters are saved permanently on the DI-100 
until you perform another millivolt calibration. You may start to use your load cell now: 
• Type: ? <Enter> to view the command list. 
•  To tare (i.e., reset the zero) the load cell, type: TARE <Enter> 
•  To view weight once, type: W <Enter> 
•  To view weight continuously, type: WC <Enter> 
•  Pull or push on the load cell, the weights should change accordingly. 
• Press <Enter> to stop the scrolling. 