8.6 Phantom
<SAM Twin Phantom>
Shell Thickness
2 ± 0.2 mm;
Center ear point: 6 ± 0.2 mm
Fig. 8.7 Photo of SAM Twin Phantom
Filling Volume
Dimensions
Approx. 25 liters
Length: 1000mm; Width: 500mm;
Height: adjustable feet
Measurement
Areas
Left Head, Right Head, Flat phantom
The bottom plate contains three pair of bolts for locking the device holder. The device holder positions are
adjusted to the standard measurement positions in the three sections. A white cover is provided to tap the
phantom during off-periods to prevent water evaporation and changes in the liquid parameters. On the phantom
top, three reference markers are provided to identify the phantom position with respect to the robot.
<ELI4 Phantom >
The ELI4 phantom is intended for compliance testing of handheld and body-mounted wireless devices in the
frequency range of 30MHz to 6 GHz. ELI4 is fully compatible with the latest draft of the standard IEC 62209-2
and all known tissue simulating liquids.
ELI4 has been optimized regarding its performance and can be integrated into a SPEAG standard phantom
table. A cover prevents evaporation of the liquid. Reference markings on the phantom allow installation of the
complete setup, including all predefined phantom positions and measurement grids, by teaching three points
The phantom can be used with the following tissue simulating liquids:
Water-sugar based liquids can be left permanently in the phantom. Always cover the liquid if the system is
not in use; otherwise the parameters will change due to water evaporation.
DGBE based liquids should be used with care. As DGBE is a softener for most plastics, the liquid should
be taken out of the phantom and the phantom should be dried when the system is not in use (desirable at
least once a week).
Do not use other organic solvents without previously testing the phantom resistiveness.
Fig.8.8 Photo of ELI4 Phantom