Central Station Items of Interest:
Number of call attempts: 9 times - alternating between the
rstand
second
numbers programmed.
Supervision of Pendants Yes – all pendants check in with the Base every 13 hours. A Pe
ndant
mi
ssing for
more
tha
n 52 hours signals a “Loss of RF” condition/report.
Reporting of Low Pendant Battery to Central Station
Upon receipt of
rst
“Replace Battery Now” Reporting of AC Power Loss After 8 hours
of power loss - and only once. Reporting of Restore Codes Only once after restore
condition met.
(Dealer programmable feature).
Operator Commands It can communicate in Duplex of Simplex mode. Call starts in Duplex but can be
switched to Simplex at any time by pushing 1 to talk and 3 to Listen
Unlimited talk time; but System will hang-up after 3 minutes of silence.
“4” will switch back to duplex mode
“7” will extend time another 3 minutes
“9” will force a hang-up and reset
Pendant battery life 5 year service life - estimated
ContactIDEventCodeProle: StandardAdemcoContactID
Contact ID Event Codes Sent:
1-101 Personal Emergency (personal emergency)
1-606 Listen-in to follow (request for voice call)
1-301 AC Loss (AC power loss longer than 8 hours)
3-301 Restoral - AC Loss
1-381 Loss of Radio supervision (pendant has not reported to the base in the last 52 hours, four 13 hour cycles)
3-381 Restoral - Loss of Radio supervision
1-384 RF low battery (pendant batter is < 3.5 volts)
3-384 Restoral - RF low battery
1-602
Periodic
test
report (periodic intervals
are
installer programmable
to 1, 7 or 30 days - default is 7 days)
Note: Each pendant is set up as a
Zone -
if
a
pendant event code
is
sent, the central s
tation
can know w
hich
pendant
initiated the call
(only if the
installer
notiesthecentralstationwhichpersoncarrieseachpendant).
4+2EventCodesProle:
01, 02, 03, 04
Medical
Alarms, zones 1, 2, 3, 4
40 A/C Fail
50 A/C Restoral
61, 62, 63, 64 R/F Loss of Supervision, zones 1, 2, 3, 4
71, 72, 73, 74 R/F Restoral of Supervision, zones 1, 2, 3, 4
41, 42, 43, 44 R/F Low Battery, zones 1, 2, 3, 4
51, 52, 53, 54 R/F Battery Restoral, zones 1, 2, 3, 4
30 Timer Test
Note: Each pendant & EWC is treated as a separate zone