QUICK GUIDE | ShuttleSoft 3
3
8
8.1
5
4 4.1
LOLIGO
®
SYSTEMS1.1
Download the latest version of ShuttleSoft 3 from our website: loligosystems.com/downloads
Follow the installation instructions on the screen and then restart the PC.
For troubleshooting, please visit our online FAQ:
loligosystems.com/faq
1
Connect the green (WiBu) copy protection dongle to a USB port on your PC (3a).
3
Connect the Long range Bluetooth dongle to a USB port on your PC, and let it initialize (3b).
Disable any other Bluetooth radios on your PC.
Connect the USB C camera to a USB C (or >USB 3.0) port on your PC, and let it initialize (4).
The LED on the back of the camera will light green, when the camera is ready to use. Keep the
spacer ring (green arrow, 4) between the camera house and lens to get correct focus.
Connect the PowerX4 power strip to a grounded wall outlet (4.1a).
Connect the WTW meter to a USB port on your PC. Power on the meter.
If you are using a Witrox oxygen instrument... Connect the Witrox 1/4 to a power source (PC
USB or USB adapter) (4.1b/c). Click the red Power button. The blue connect icon will now ash
blue (for 5 min.) indicating that the Witrox is in pairing mode.
4
FIRST TIME USE
LIVE EXPERIMENTS
Power on the WTW meter by pressing the Power button. Press and hold the ENTER button to
enter the Storage & cong menu. Select System > Interface > Baud rate, and set the baud rate
to 19200. Press the F1 button until you return to the start screen (3c).
2
Open ShuttleSoft 3 > Devices > Choose Scan for new devices. When scanning is complete,
press and hold the F2 button on the WTW meter until the Autom. USB Output menu appears.
Change Interval rate, if necessary. Otherwise, choose Continue to exit.
Go back to the main menu, when all connected instruments have been found (5).
5
Repeat step 3-4, and open ShuttleSoft 3. Now choose Use current conguration and
complete step 5.
6
Video tab: Shows the live feed from the camera. Any mask, zones or lter can be turned on/o
at the bottom of the screen.
9
Main menu > Live experiments: The software will now load available cameras; one for
each experiment tab. Close any unwanted cameras. Click on the Experiment tab to open the
experiment (green arrow, 8).
Start by performing a pixel-calibration to enable correct distance measurements. Drag the
blue ruler endpoints to dene a known length (e.g., by placering a ruler in the background)
on the live preview (8.1). Input the dened length in the Calibration length eld. Adjust the
resolution and area of interest (AOI) to improve performance. Click Accept settings to continue.
IMPORTANT: Once the camera is in place, the camera must not be moved! Otherwise, a new
pixel-calibration is necessary to continue.
8
The WTW oxygen sensor must be calibrated with the WTW meter. See the WTW meter user
manual for more instructions.
If you are using a Witrox oxygen instrument... Main menu > Calibration > Choose Witrox
instrument: Each oxygen sensor must be calibrated before use. Follow the Quick guide for
Witrox Instrument to perform a manual calibration. The temp probe is pre-calibrated.
7
a
c
a
b c
b