Series IN 500 Operating Manual Data format UPP (Universal Pyrometer Protocol) 33
Reading hysteresis:
AAh
l
XX
XX (2
-
digit hex)
2)
;
XX setting 2 ... 20°C (4 ... 36°F)
Setting hysteresis:
AAh
l
Output: XX (2
-
digit hex)
2)
Wait time:
AAtwXX
XX
= 00 ... 99 relative delay value
Reading sensor data:
AAse
Output: XXXXYYYY (8
-
digit decimal); Reading
of adjusted sensor data: XXXX: S1; YYYY: S2
Setting sensor data:
AAseXXXXYYYY
XXXXYYYY (8
-
digit dec.)
; entering sensor data
Reset
1)
:
AAre
Reset
device
Reading parameters:
AApa
Output 11
-
digit decimal:
Digit 1 and 2 (10...99 or 00): EMI up to 100
Digit 3 (0 ... 6): t
90
(response time)
Digit 4 (0 ... 8): t
CL
(storage clear mode)
Digit 5 (0 / 4): analog output
Digit 6 and 7 (00 ... 99): sensor head temp.
Digit 8 and 9 (00 ... 31): device address
Digit 10 (0 ... 4): device baud rate
Digit 11 (always 0)
Error status:
AAfs
Output:
XX;
XX=00…FF (00 = no error)
(01…FF: error code for LumaSense service)
Serial number:
AASsn
Output: XXXXX (5
-
d
igit decimal)
Type / software
version:
AAve
Output: XXYYZZ (6
-
digit decimal)
XX = 76 (IN 510, IN 520, IN 530);
YY = Month of software version
ZZ = Year of software version
Notes: The letter “l” means the lower case letter of “L”
Bold values = default settings
After entering these commands the device carries out an automatic reset. The device
needs approx. 150 ms before it is ready to use and work with the changed settings.
The input and output corresponds to the preset degree C or degree F.
Additional instruction for the RS485 interface
Requirements to the master system during half-duplex operation:
1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some
older interfaces are not fast enough for this).
2. The pyrometer's response will follow after 5 ms at the latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be
repeated.
After receiving the response, the master has to wait at least 1.5 ms before a new command can
be entered.