Owner's manual

Chapter 4
_________________________________________________________________________________________
_____________________________________________________________________________
BE6024-12 1316A-3 Multi Gas Monitor LumaSense Technologies A/S
Instruction Manual Page 69 of 96
Figure 4.14 The Status window displaying that a new O
2
sensor is required.
Prior to any span calibration perform a Zero/O
2
span calibration after
completion of the warm up time.
Initially set Zero Calibration to "Manual" in “Zero Group”, see figure 4.15.
This will avoid the Zero calibration to interfere with the Span calibration.
Figure 4.15 Set Zero calibration to Manual
Start the measurement by pressing the “Start”-button in the Main window, see
figure 4.16