5.1 Maintenance Items
5.1.1 Changing Zero Channel inlet filter
It is recommended to change the zero gas inlet filter at least once a
year.
- Remove the top cover.
- Remove the DS0790 filter by detaching the luerlok fittings.
- Install a new DS0790 filter.
Figure 5.1. Picture showing the DS0790 zero inlet dust filter.
5.1.2 Changing the fuses
On delivery, two 20 mm T 1.6A (VF0007A) slow-blow fuses are
installed in the monitor one in the live and one in the neutral mains
feed. When replacing fuses always ensure that the correct fuses are
used.
To check and change the fuses:
1. Remove the power cable from the mains socket.
2. Insert a suitable screwdriver into both of the fuse holders and
unscrew the fuse holders. See figure 5.2.