Luminex® LDS 1.7 SP1 Upgrade Instructions to xPONENT® 3.
© LUMINEX CORPORATION, 2001-2009. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language or computer language, in any form or by any means without prior express, written consent of: LUMINEX CORPORATION 12212 Technology Boulevard Austin, Texas 78727-6115 U.S.A. Voice: (512) 219-8020 Fax: (512) 219-5195 Luminex® LDS 1.7 SP1 Upgrade Instructions to xPONENT® 3.1 SP1 PN 89-00002-00-211 Rev.
Contents Upgrade Procedure .............................................................................1 Important Upgrade Information..........................................................2 Recommendations...............................................................................2 Training...............................................................................................2 System Requirements .........................................................................2 Kit Components ......
ii
Upgrade Procedure 1
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use Important Upgrade Information This Luminex xPONENT 3.1 SP1 product conversion kit is intended for use on a single Luminex analyzer (Luminex 100 or Luminex 200™) purchased after October 1, 2000. If you install this kit on multiple analyzers, you accept the legal and financial risk associated with violating software licensing.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use • • Kit Components Note: For information about calibration and verification frequency and procedures, see Luminex xPONENT 3.1 Software Manual. 2000. Contact Luminex Technical Support with the analyzer serial number to confirm that it meets the requirements. A Luminex XYP purchased after May 30, 2001.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use These license packages are part of the upgrade kit only if you purchase them from Luminex: • 21 CFR Part 11 package • LIS package • Automation package • Security package • Extra seats package Computer and Software Upgrade Instructions Install New PC Note: If you did not get a new PC with your xPONENT 3.1 SP1 upgrade, skip this step. You must run Luminex xPONENT 3.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use Archive “My Sessions” Folder Note: You will not need to archive the “My Sessions” folder if you are upgrading your PC. Uninstall Luminex LMAT Software Note: You will not need to uninstall Luminex software if you are upgrading your PC. You cannot import data into xPONENT 3.1 SP1 from LDS1.7 SP1; however, you can store the data to import to another LDS1.7 system at a later time.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use 3. Double-click the Add/Remove Programs icon. The Add/Remove Programs dialog box opens. 4. Under the Currently installed programs list, select the Luminex Data Collector entry. The entry expands to include the Change/Remove button. Click Change/Remove. 5. The InstallShield Wizard dialog box opens. Select Remove and click Next. 6. Click OK in the Confirmation File Deletion dialog box.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use 9. A dialog box opens with two installation options, Complete and Custom. Select Complete and click Next to continue. 10. Click Install at the next dialog box to continue installation, which may take several minutes. Note: Please see “Licensing” in the “Using” chapter of the xPONENT 3.1 software manual if you need information on how to change licensing information. Note: The xPONENT 3.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use If you are using an existing computer, perform the Luminex software update before upgrading the firmware. See "Install Luminex xPONENT 3.1 SP1 Software". Luminex SD System Firmware Upgrade To upgrade the Luminex SD system firmware: 1. Close any open applications. 2. Turn off the power to the SD, XYP, and analyzer. 3.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use 6. Select the SD Device option button (item 6, above). Wait until the top line of the Luminex XYP/SD Instrument Firmware Version text box (item 4, above) displays the current version of the Luminex SD system firmware and the second line displays its status. This may take up to ten minutes. The example above may not match the dialog box on your system. 7.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use LX100 Sample Probe If you have a Luminex 200 (you can tell that it’s a Luminex 200 if it has the sample probe shown in the next photo), click Upgrade adjacent to the Detector Device text box. The Confirm Firmware Download dialog box opens. Click Yes to continue with the upgrade. LX200 Sample Probe 5. The Download In Progress dialog box opens displaying the progress. Read the information presented on the dialog box.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use plugging the serial end into P1 on the analyzer, and the USB end into an available USB port on the computer. 8. Connect the short CANBUS cable (gray) to P1 on the back of the SD and P2 on the back of the LX100 or LX200 analyzer. 9. Turn the power on to the Luminex analyzer. 10. Turn on the power to the SD. 11. An entry in the Detector Device text box will display the new firmware version.
Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use Obtain License Key You must obtain a license key from Luminex Technical Support to enable the full base package beyond the 90-day trial period. In order to assign your license key, we will need information about your Luminex system. Please have the following information available: • The serial number of your Luminex system. • The serial number of your Luminex XYP.