Hardware and Installation Manual Owner's manual

Installation Category II
Pollution degree 2
Temperature control: maintains samples using the heater block at a constant temperature
from 35° C to 60° C (95° F to 131° F) +/- 1° C of set point.
Automatic transfer of assay protocols and new reagent information into the system using a
large capacity read/write DVD
Automatic sampling from a 96-well microtiter plate, beginning from any well position
Automatic real-time analysis
Analysis of multiple assay protocols per microtiter plate
Barcode reader entry of sample IDs
Produces sound pressure levels below 85 dBA
Environmental Conditions
Indoor use only
Operating temperature: 15-35°C (59-95°F)
Operating relative humidity: 20-80%, noncondensing
Operating altitude: up to 2400m (7874 ft) above mean sea level
Shipping temperature: 0-50°C (32-122°F)
Storage temperature: 10-40°C (50-104°F)
Electronics
USB 2.0-compatible communications link for fast data transfer
Input voltage range: either 100-120 V~, 2.0 A 50/60 Hz or 200-240 V~, 1.0 A 50/60 Hz
Optics
Reporter channel detection: A/D resolution 16 bits
Reporter detector: CCD, detection bandwidth of 566 to 614 nm
Classification detector: CCD
Limit of Detection (LOD): For the reporter channel, using a blank microsphere from region
078, ≤ 700 molecules of phycoerythrin (PE) per microsphere
Dynamic Range: For the reporter channel, using a microsphere from region 078, ≥3.0
decades
Efficiency: Classification Channels: ≥ 80%
Fluidics
Sample load rate: 20 µL to 500 µL per second
Sample uptake volume: 20 to 200 mL
Well-to-well carryover: ≤ 4%
Sample uptake accuracy: +/- 5%