WRIST TYPE DIGITAL BLOOD PRESSURE MONITOR MODEL 1143 INSTRUCTION MANUAL Important! Read and understand these instructions before using the Lumiscope Blood Pressure Monitor. If you do not understand any part of these instructions, contact your medical professional or GF dealer for direction in the use of this product. If components are damaged or missing, contact your GF dealer immediately. DO NOT use substitute parts. GF Health Products, Inc.
CONTENTS PARTS AND COMPONENTS.............................................................................................................3 SYMBOLS..........................................................................................................................................4 1 GENERAL....................................................................................................................................5 PRINCIPLE OF OPERATION...........................................................
PARTS AND COMPONENTS 1. Main Body 2. Display 3. Wrist Cuff 4. Button ‘O/I’ 5. Button ‘M’ 6. Batteries 7.
SYMBOLS Symbol Meaning Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC. The device, accessories and packing must be disposed of as waste correctly at the end of the usage. Please follow Local Ordinances or Regulations for disposal.
1 GENERAL This instruction manual is intended to assist the user in safe and efficient operation of the wrist type digital blood pressure monitor (hereinafter: device) model 1143. The device must be used in accordance with the procedures described in the manual. It is important to read and understand the entire manual, especially the section .
Note: This device cannot provide reasonable accuracy if used or stored at temperature or humidity beyond the range stated in the section of this manual.
2 TIPS ON TAKING BLOOD PRESSURE MEASUREMENT 1. It is necessary to know that arterial blood pressure is subjected to sharp fluctuations. The level of the arterial blood pressure depends on many factors. Generally arterial blood pressure is lower in the summer and higher in the winter. Arterial blood pressure changes with atmosphere pressure and is affected considerably by many factors, e.g. physical loads, emotional excitability, stress, meals, etc.
The illustration above is from The British Hypertension Society 2. Patients with cardiovascular diseases: Take measurement under doctor’s instruction. Under no circumstances should you alter the dosage of any drug prescribed by your doctor! 3. Accurate measurement of blood pressure may be difficult in people with serious arteriosclerosis, weak pulse, or in patients with irregular heart rates. Please consult a qualified physician to interpret your blood pressure readings.
4. Keep quiet and relax during measurement. To obtain accurate readings, wait one hour after a meal or smoking before taking a measurement. 5. Wait at least 5 minutes before taking the next measurement. Atherosclerosis and diabetes patients require longer intervals (10-15 minutes), since vessel elasticity may have decreased in these pathologies.
3 BATTERY INSTALLATION 1. Open the battery cover and install two alkaline AAA batteries in the battery compartment as indicated. Check polarity and ensure the batteries are correctly positioned; 2. Close the battery compartment cover. • Replace the batteries when the replacement indicator appears in the display or when the unit does not turn on when button ‘O/I’ is pressed; • Use only alkaline AAA batteries, do not use rechargeable batteries; • Use only same-age, same-type batteries together.
4 CORRECT POSTURE FOR MEASUREMENT Sit down and place your elbow on a table. Ensure the wrist is at approximately the same level as the heart, and the forearm is extended naturally, palm up. Wrap the cuff around the left wrist as shown in the illustration, and ensure the cuff is attached to the wrist securely with the display facing the user. Loosen and roll up sleeves so that the blood pressure monitor wrist cuff is in direct contact with the wrist during measurement.
5 SETTING THE DATE AND TIME The device has a 12-hour time format. The icon ‘AM’ indicates the time of Ante Meridiem and the icon ‘PM’ indicates the time of Post Meridiem. To get accurate date and time, the user should preset the date and time correctly before the first use of this device. The operation procedure for presetting Date/Time is as follows: 1. When the device has batteries installed, the display will show as in Fig. 1; Fig.1 2.
3. Press button ‘M’ again to increase the number, and press button ‘O/I’ for confirmation; 4. When the year setup is confirmed, the month number will flash automatically. Follow the same instruction as above to set month, day and time; Fig.3 5. Press button ‘O/I’ to finish setup. If you want to change the date and time, repeat steps 2, 3, and 4.
6 TAKING A MEASUREMENT 1. Before taking a measurement, take 3 ~ 5 deep breaths and relax. Do not talk or move your arm during measurement; 2. Press button ‘O/I’, all symbols will display for 2 seconds as in Fig.4; then ‘0’ will display. The pump will then begin to inflate while displaying pressure. Generally the pressure will reach 190 mmHg as in Fig.5; 3. The pump stops inflating and pressure begins to decrease gradually, during which the user’s blood pressure will display as in Fig.
5. Press button ‘O/I’ to return to standby mode. Wait for at least 5 minutes before taking another measurement. If the device remains unused for 3 minutes, it will return to standby mode automatically. AUTOMATIC INFLATION This device has 4 levels of inflation pressure: 190 mmHg, 230 mmHg, 270 mmHg and 300 mmHg. When 190 mmHg is insufficient, or arm movement occurs, the device will automatically inflate to an appropriate pressure level in order to obtain the blood pressure measurement.
7 MEMORY FUNCTION MEMORY RECALL 1. Model 1143 can store 90 sets of readings and will automatically calculate the average of the last 3 readings. When memory is full (90 sets of readings are stored), the oldest reading will be replaced by the newest. Memory will not clear even if the batteries are removed; 2. After a measurement is finished, or when the device is in standby mode, the user can press button ‘M’ to recall memory. Press button ‘M’ to display the average value of the last 3 readings as in Fig.
3. Press button ‘M’ again, the device displays ‘01’; then after 1-2 seconds the latest reading will display as in Fig.9; Fig.9 4. Press button ‘M’ again, the device displays ‘02’; then after 1-2 seconds the second-to-the-last reading will display. DELETING THE MEMORY To clear memory: After a measurement is finished or when the device is in standby mode, press and hold button ‘M’ for at least 5 seconds. The device will display ‘Clr’, which means the stored reading is deleted as in Fig.10. Fig.
8 WHO BLOOD PRESSURE CLASSIFICATION The World Health Organization (WHO) has established standards for assessment of high or low blood pressure, regardless of age, as shown in the chart below: This device has an indicator located on the display which displays a segment, based on the current data, corresponding to the WHO classification.
Note: If the systolic blood pressure and diastolic blood pressure fall into different categories, use the higher value for classification.
9 ERROR AND LOW BATTERY INFORMATION INDICATION POSSIBLE REASON CORRECTION METHODS The wrist cuff is positioned incorrectly Ensure wrist cuff is correctly positioned; repeat measurement Movement of arm (or hand) or talking during measurement Repeat measurement, following recommendations of this manual The wrist cuff is not inflated to sufficient pressure Repeat measurement Arrhythmia Consult your physician The batteries are weak Replace both batteries with new ones 20
10 CARE, STORAGE, REPAIR AND RECYCLING 1. It is necessary to protect this device against high moisture, direct sunlight, shock, alcohol and gasoline. 2. Remove the batteries if the device is to be stored for a long period of time, and keep the batteries away from children. 3. Prevent sharp objects from coming into contact with the wrist cuff, and do not stretch or twist the wrist cuff. 4. Use only clean, soft, dry cloths to clean the device. 5. The wrist cuff is sensitive and must be handled with care.
11 TROUBLESHOOTING SYMPTOM CHECK POINT REMEDY No display when button ‘O/I’ is pressed The batteries need to be replaced Replace both batteries with new ones The batteries are incorrectly positioned The battery contacts are corroded Inflation stops and re-inflates later Install the batteries in correct position Clean the battery terminals with a dry cloth Inflation is automatic to ensure correct measurement See Did you talk or move your arm (or hand) during measurement? Keep
Troubleshooting continued The reading is extremely low or high Pulse rate is too low or too high The batteries wear out too quickly Is the wrist cuff at the same level as the heart? Ensure your posture is correct Is the wrist cuff wrapped correctly? Wrap the wrist cuff correctly Did you strain your arm (or hand) during measurement? Relax during measurement Did you talk or move your arm (or hand) during measurement? Keep quiet and silent during the measurement Did you talk or move your arm (or han
12 SPECIFICATIONS Model 1143 Size (L x W x H) 2 3/4" x 2 3/8" x 1 1/4" (70 mm x 60 mm x 30 mm) Weight (without batteries) ~ .24 lb (110 g) Measuring method Oscillometry Measuring range 40 to 260 mmHg (blood pressure) 40 to 160 beats/minute (pulse rate) Measuring accuracy ± 3 mmHg for systolic and diastolic pressure ± 5% of the reading for the pulse rate Inflation Automatic pump Rapid deflation Automatic electronic valve Batteries 2 alkaline AAA x 1.
Specifications continued Memory 90 sets of measurements Temperature and humidity Operation: 50°F to 104°F (10℃ to 40℃), RH 85% and below Storage: -4°F to 122°F (-20℃ to 50℃), RH 85% and below Cuff size Fits wrist size 5" - 8" (127 mm - 203 mm ) Complete kit Main body, wrist cuff, storage case, 2 alkaline AAA batteries, instruction manual, gift box 25
13 TWO-YEAR LIMITED WARRANTY GF Health Products, Inc. guarantees this product free from defects in material and workmanship for a period of two years from the date of purchase, except as noted below: This product warranty does not cover damage caused by misuse or abuse, the attachment of any unauthorized accessory, alteration to the product, or any other conditions whatsoever that are beyond the control of GF Health Products, Inc. GF Health Products, Inc.
Warranty is solely through the Repair Department. Service of this product by anyone other than the Repair Department voids warranty. Graham-Field® and Lumiscope® are registered trademarks of GF Health Products, Inc. Packaging, warranties, products, and specifications are subject to change without notice. GF Health Products, Inc. is not responsible for typographical errors.
GF Health Products, Inc. 2935 Northeast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 © October 2009 GF Health Products, Inc. Made in China www.grahamfield.