6000 W x 1 W, PF, ACV, ACA WATT METER Model : DW-6163 Your purchase of this WATT METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and delicate instrument, its durable structure will allow many years of use if proper operating techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach.
Caution Symbol Caution : * Risk of electric shock ! Caution : * Do not apply the overload voltage, current to the input terminal ! * Remove test leads before open the battery cover ! * Cleaning - Only use the dry cloth to clean the plastic case ! Environment Conditions * Installation Categories III-600V. * Pollution Degree 2. * Altitude up to 2000 meters. * Indoor use. * Relative humidity 80% max.
TABLE OF CONTENTS 1. FEATURES....................................................................... 1 2. SPECIFICATIONS........................................................... 1 2-1 General specifications.............................................. 1 2-2 Electrical specifications................................................ 2 3. FRONT PANEL DESCRIPTION............................................ 4 3-1 Display........................................................................
1. FEATURES * Professional precision WATT meter with 1 W resolution digital display , battery operated. * LSI - circuit provides high reliability and durability. * Directly operation. * Measurement : WATT ( AC ) : 6000W x 1W. ACV : 600V x 0.1 V . ACA : 10 A x 10 m A . PF : 1.00 x 0.01 * True Power and Wide range 0 W to 6000 Watt. * True RMS measurement of ACV / ACA. * Super large LCD , Dual display at same time. * Full function Auto range. * Memory Record ( Max. , Min. ) , Data Hold.
Data Hold Freeze the display reading . Memory Recall Maximum & Minimum value . Data output RS232 / USB PC Computer interface. * Connect the optional RS232 cable UPCB - 02 will get the RS232 plug. * Connect the optional USB cable USB - 01 will get the USB plug. Operation 0℃ to 50℃ ( 32℉ to 122℉). Temperature Operation Less than 80% RH Humidity Power Supply Alkaline or Heavy duty type DC 9V battery 006P , MN1604 (PP3) or equivalent. Power Approx. DC 10 mA Consumption Weight 398 g/0.
V/A ( TRUE RMS ) Range Resolution Accuracy ACV 600 V ACA 10 A 0.1 V 0.01 A ± ( 0.3 % + 0.3 V ) ± ( 0.3 % + 0.03 A ) * Auto range. * ACV, ACA accuracy is test under input signal is sine wave, 50/ 60 Hz. * ACV ACA frequency response is from 40 to 400 Hz. * Max. input value, AC voltage : 600V, AC current : 10 A. * Accuracy @ 23℃ ± 5℃ PF ( Power factor ) Range Resolution Accuracy 1.00 0.
3. FRONT PANEL DESCRIPTION Fig. 1 3-1 Display 3-2 POWER / Backlight button 3-3 FUNCTION button 3-4 HOLD button 3-5 RECORD button ( MAX. / MIN.
4. MEASURING PROCEDURE 4-1 Power ON/OFF 1)Press the POWER button ( 3-2, Fig. 1 ) more than 1.5 sec to Turn ON the meter. 2)Press the POWER button ( 3-2, Fig. 1 ) more than 1.5 sec to Turn OFF themeter. 4-2 ACV measurement procedure 1)Press the POWER button ( 3-2, Fig.1 ) more than 1.5 sec to Turn ON the meter. 2)Push the FUNCTION button ( 3-3, Fig. 1 ) once then change function for measuring ACV. 3)Connect Red test lead to the " V " input terminal and Black test lead to " COM " input terminal .
4-4 Watt measurement procedure 1)Press the POWER button ( 3-2, Fig. 1 ) more than 1.5 sec to Turn ON the meter. 2)Make sure your equipment work voltage, such as 110 V or 220 V. 3)Turn off the POWER SOURCE. 4)Plug in the LOAD ( equipment ) to V terminal and COM terminal. 5)Plug in the power source to WATT terminal and 10A terminal. 6)Push the FUNCTION ( 3-3, Fig.
2)With the " REC. " symbol on the display : a) Press the " REC Button " ( 3-5, Fig. 1 ) once, the " REC. MAX. " symbol along with the maximum value will appear on the display. If intend to delete the maximum value, just press the " Hold Button " ( 3-4, Fig. 1 ) once, then the display will show the " REC. " symbol only & execute the memory function continuously. b)Press the " REC Button " ( 3-5, Fig. 1 ) again, the " REC. MIN. " symbol along with the minimum value will appear on the display.
5. MAINTENANCE 1)This WATT METER is intended for measuring the TURE RMS POWER CONSUMPTION. At the same , It can let you know .how many effect from your equipment. 2)When measuring POWER CONSUMPTION. Equipment working voltage must be recheck again, before turn on the equipment power switch. 3)Instruments used in dusty environments should be stripped and cleaned periodically. 4)Do not leave the instrument exposed to direct heat from the sun for long periods.
Meter PC (9W 'D" Connector) Center Pin..........................Pin 4 (3.5 mm jack plug) Ground/shield.......................Pin 2 2.
RS232 FORMAT : 9600, N, 8, 1 Baud rate 9600 Parity No parity Data bit no. 8 Data bits Stop bit 1 Stop bit 7. BATTERY REPLACEMENT 1)When the Top of left corner on LCD display show " ", it is necessary to replace the battery , However in -spec. measurement may still be made for several hours after LOW BATTERY INDICATOR appears before the instrument become inaccurate. 2)Open the screw of " Battery Cover " by screwdriver, then move the battery. 3)Replace with 9V battery and reinstate the cover. 8.