Datasheet

SD Card real time data logger
4 channels THERMOMETER
Model : TM-947SD
FEATURES
* Type K/J/T/E/R/S, Pt 100 ohm, measurement with 4 display. Power of
f
Auto shut off saves battery life or
* Show 4 channels display on the LCD at the same time. manual off by push button.
* Type K : -100 to 1300 . Operating 0 to 50 .
* Type J : -100 to 1200 . Temperature
* Pt 100 ohm : -199.9 to 850.0 . Operating Less than 85% R.H.
* / , 0.1 degree/1 degree.℃℉ Humidity
* 4 channels ( T1, T2, T3, T4 ), T1-T2. Power Supply
*
A
Alkaline or heavy duty DC 1.5 V battery
* Microcomputer circuit provides intelligent function ( UM3, AA ) x 6 PCs, or equivalent.
and high accuracy.
*
A
DC 9V adapter input. ( AC/DC power
* Offset adjustment for the Type K/J/T/E/R/S measurement. adapter is optional ).
* Offset adjustment for the Pt 100 measurement. Power Current Normal operation ( w/o SD card save
* Measuring unit can select to or .℃℉ data and LCD Backlight is OFF) :
* Real time SD memory card Datalogger, it Built-in Clock
Approx. DC 8.5 mA.
and Calendar, real time data recorder, sampling time set When SD card save the data but and
from 1 second to 3600 seconds. LCD Backlight is OFF) :
* Manual datalogger is available ( set the sampling
Approx. DC 30 mA.
time to 0 second ), during execute the manual datalogger
*
A
If LCD backlight on, the powe
r
function, it can set the different position ( location ) No.
consumption will increase approx.
( position 1 to position 99 ).
14 mA.
* Innovation and easy operation, computer is not need to Weight 489 g/1.08 LB.
setup extra software, after execute datalogger, just take Dimension 177 x 68 x 45 mm
away the SD card from the meter and plug in the SD card (7.0 x 2.7x 1.9 inch)
into the computer, it can down load the all the measured Accessories * Instruction manual...........................1 PC
value with the time information ( year/month/date/ Included
hour/minute/second ) to the Excel directly, thenuser can Optional * Type K thermocouple probe.
make the further data or graphic analysis by themselves. Accessories TP-01, TP-02A. TP-03, TP-04
* SD card capacity : 1 GB to 16 GB. * Pt 100 ohm probe, TP-101.
* LCD with green light backlight, easy reading. *SD Card ( 1 GB )
* Can default auto power off or manual power off. *SD Card ( 2 GB )
* Data hold, record max. and min. reading. * USB cable, USB-01.
* Microcomputer circuit, high accuracy. * RS232 cable, UPCB-02.
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter. * Data Acquisition software,
* RS232/USB PC COMPUTER interface. SW-U801-WIN.
* Heavy duty & compact housing case. * AC to DC 9V adapter.
* Hard carrying case, CA-06.
GENERAL SPECIFICATIONS * Soft carrying case, CA-05A.
Circuit Custom one-chip of microprocessor LSI
circuit. ELECTRICAL SPECIFICATIONS (23± 5 )
Display LCD size : 52 mm x 38 mm
LCD with green backlight ( ON/OFF ). PT 100 ohm
Channels T1, T2, T3, T4, T1-T2. Resolution Range Accuracy
Sensor type Type K thermocouple probe. 0.1 -199.9 to 850.0 ± ( 0.4 % + 1 )
Type J/T/E/R/S thermocouple probe. 0.1 -327.0 to 999.9 ± ( 0.4 % + 1.8 )
PT 100 ohm probe 1 1000 to 1562 ± ( 0.4 % + 2 )
*
Cooperate with an 0.00385 alpha
*
Pt 100 ohm probe TP-101 is the optional accessory.
coefficient, meet DIN IEC 751.
Resolution 0.1/1, 0.1/1 .℃℃ Type K/J/T/E/R/S
Datalogger Auto 1 second to 3600 seconds Sensor Resolution Range Accuracy
Sampling Time
@ Sampling time can set to 1 second,
Type
Setting range
but memory data may loss.
Type K 0.1 -50.1 to -100.0 ± ( 0.4 % + 1 )
Manual Push the data logger button -50.0 to 999.9 ± ( 0.4 % + 0.5 )
once will save data one time. 1 1000 to 1300 ± ( 0.4 % + 1 )
@ Set the sampling time to
0.1 -58.1 to -148.0 ± ( 0.4 % + 1.8 )
0 second.
-58.0 to 999.9 ± ( 0.4 % + 1 )
@ Manual mode, can also select the
1 1000 to 2372 ± ( 0.4 % + 2 )
1 to 99 position ( Location ) no.
Type J 0.1 -50.1 to -100.0 ± ( 0.4 % + 1 )
Memory Card SD memory card. 1 GB to 16 GB. -50.0 to 999.9 ± ( 0.4 % + 0.5 )
Advanced * Set clock time ( Year/Month/Date, 1 1000 to 1150 ± ( 0.4 % + 1 )
setting Hour/Minute/ Second ) 0.1 -58.1 to -148.0 ± ( 0.4 % + 1.8 )
* Decimal point of SD card setting -58.0 to 999.9 ± ( 0.4 % + 1 )
* Auto power OFF management 1 1000 to 2102 ± ( 0.4 % + 2 )
* Set beep Sound ON/OFF Type T 0.1 -50.1 to -100.0 ± ( 0.4 % + 1 )
* Set temperature unit to or ℃℉ -50.0 to 400.0 ± ( 0.4 % + 0.5 )
* Set sampling time 0.1 -58.1 to -148.0 ± ( 0.4 % + 1.8 )
* SD memory card Format -58.0 to 752.0 ± ( 0.4 % + 1 )
Temperature Automatic temp. compensation for the Type E 0.1 -50.1 to -100.0 ± ( 0.4 % + 1 )
Compensation type K/J/T/E/R/S thermometer. -50.0 to 900.0 ± ( 0.4 % + 0.5 )
Linear Linear Compensation for the full range. 0.1 -58.1 to -148.0 ± ( 0.4 % + 1.8 )
Compensation -58.0 to 999.9 ± ( 0.4 % + 1 )
Offset Available for Type K/J/T/E/R/S and 1 1000 to 1652 ± ( 0.4 % + 2 )
Adjustment Pt 100 ohm. Type R 1 0 to 600 ± ( 0.5 % + 1 )
Probe Input Type K/J/T/E/R/S 601 to 1700 ± ( 0.5 % + 1 )
Socket 2 pin thermocouple socket. 1 32 to 1112 ± ( 0.5 % + 2 )
Pt 100 ohm : Ear phone socket. 1113 to 3092 ± ( 0.5 % + 2 )
Over Indication Show " - - - - ". Type S 1 0 to 600 ± ( 0.5 % + 1 )
Data Hold Freeze the display reading. 601 to 1500 ± ( 0.5 % + 1 )
Memory Recall Maximum & Minimum value. 1 32 to 1112 ± ( 0.5 % + 2 )
Sampling Time Approx. 1 second. 1113 to 2732 ± ( 0.5 % + 2 )
of Display
Remark :
Data Output RS 232/USB PC computer interface.
a. Accuracy value is specified for the meter only.
* Connect the optional RS232 cable
b. Accuracy is tested under the meter's environment temperature
UPCB-02 will get the RS232 plug.
within 23 ± 5 .
* Connect the optional USB cable
c. Linearity Correction :
USB-01 will get the USB plug.
Memorize the thermocouple's curve into the intelligent CPU
circuit,
PATENT CHINA : ZL 2008 2 0189918.5 ZL 2008 2 0189917.0 TAIWAN : M 358970 M 359043
Germany : Nr. 20 2008 016 337.4 JAPAN : 3151214 U.S.A. : Pending
* Appearance and specifications listed in this brochure are subject to change without notice. 0909-TM947SD