6.2 Linearity calibration (for scale technician
or factory use)
Turn on the weighing machine and then turn it off.
Press and hold the TARE key and then turn on the
machine. The display will show LINE on the first row,
CAL-0 (flashing) on the second line and AD value on
the third line.
When the AD value is stable, press ZERO to calibrate
ZERO. After 2 or 3 seconds, the second line will
show 200. (ex. SCT 600)
Place a 200g weight on the platter and press ZERO
when the stable A/D value is displayed. After 2 or 3
seconds, the display will show 400.
Place a 400g weight on the platter and press ZERO
when the stable A/D value is displayed. After 2 or 3
seconds, the display will show 600.
Place a 600g weight on the platter and press ZERO
when the stable A/D value is displayed. After 2 or 3
seconds, the display will show 00000. Calibration is
now complete.
Turn off the power, then turn it back on again. Place a
weight on the platter to ensure weighing is correct. If
not, repeat the above steps.