User's Manual

Upgrading the Firmware
D-4 Reader Software Upgrade Procedure
Restoring Parameters Follow these steps:
1. Verify whether parameters were set to default during the upgrade procedure, by checking their
values. If yes, then reset their values to the saved values (see Saving Parameters on Page D-2).
Note a) Reader firmware ignores leading zeros for integers.
b) If the Reader was already configured on the site, you have to restore all application parameters
queried during the Saving Parameters step, or restart the configuration process as described in
Chapter 5, Configuration and Operation.
3. Check the values of the new parameters, if any, that were introduced by this software version.
Setting the Readers
Network Configuration
To set the Reader’s network configuration, do the following:
1. Set the Readers address to its initial value (see step 1 in Setting the Reader’s Address): by
typing:
IAD=r<Enter>
2. Store the Readers parameters, by typing:
:CONFIG:STORE<Enter>
3. Reset the Reader, by typing:
:RESET<Enter>
4. Verify that the Reader is functioning properly, by typing:
@r IVN<Enter>
where r is the Reader’s address. The Reader must respond with its software version number.
5. Verify that the parameters set in the following sections are correct:
Steps 2 and 3 from the section, Restoring Parameters above.
Step 1 in this section.