Operator's manual Manual

14.1a ME Gas sampling and gas conditioning technology 21
The CSS-VC.. is to be supplied in voltages 230V/50Hz or 115V/50-60Hz (for the circuit diagram see
Appendix). Protection is provided as standard with two 4A fuses. These are located in the cold-device
socket.
The electrical connection is via the 2m long power cord with cold-device connector depending on the
mounting of the connection plate either at the back or in the lid of the housing (see Section 12.2).
12.4.1 GROUP CONNECTION ALARM (ART. NO. 01G6170)
The electrical connection of the general alarm (condenser, fluid and flow) is via a 6-pin connector,
depending on the mounting of the connection plate, either at the back or in the lid of the housing.
The corresponding 6-pin socket is included.
The connector is wired as follows:
Figure 6 Plug assignment for the version with group alarm (Art. No. 01G6170)
Figure 7 Upper and lower channel of CSS-VC2
1
2
3
4
5
6
Internal
External
Status Alarm
Contact load
Contact rating
24V AC/DC 0,5A
Plug/male connector type Phoenix MSTB
2,5/6-STF 5,08
Socket/Female X2
Upper Channel
Channel 2
Lower channel
Channel 1