Operator's manual

18
Gas sampling and gas conditioning technology 6-1.3.5-ME
NOTE!
If none of the specified faults can be located, despite the pump not
operating correctly, the pump should be returned to M&C for examination.
When sending diaphragm pumps for repair to M&C Customer Service,
information about the transported medium is required.
Our workshop should be informed in particular about aggressive media.
Where pumps have been used for transporting dangerous or highly
aggressive gases, these should be cleaned prior to being returned.
17 SPARE PARTS
Wear, tear and replacement part requirements depend on specific operating conditions.
The recommended quantities are based on experience and are not binding.
Diaphragm Pump
Type MP30
(C) consumable parts, (S) spare parts
Recommended quantity
MP30 being in operation
(f.d. = for demand)
V/E/T 1 2 3
90P1500
Diaphragm 8
C 1 2 3
90P1510
Set sealing ring 3 (2 pc., Kalrez
)
C 1 2 3
90P1505
Set valve plate 2 (2 pc.)
C 2 4 6
PVDF male connectors with G-thread (ISO 1010031)
05V1045 Straight connector
DN 4/6-G1/8” material: PVDF
S - - 2
05V1050 Straight connector
DN 6/8-G1/8” material: PVDF
S - - 2
05V6600 Ferrule DN 4/6 PVDF S 2 2 4
05V6602 Ferrule DN 6/8 PVDF S 2 2 4
05V6605 Union nut DN 4/6 PVDF S 2 2 4
05V6607 Union nut DN 6/8 PVDF S 2 2 4
18 APPENDIX
More product documentation is available on our Internet catalogue:
www.mc-techgroup.com