Operator's manual Owner's manual

18
Gas sampling and gas conditioning technology 10-5.1-ME
The top part (1) of the door hinge can, after releasing
the 2 screws (2) at the top or bottom of the door, be
moved slightly in the direction of the arrow.
The door can be adjusted after releasing the socket
screw (3) and rotating the eccentric (4) by means of
a screwdriver. NOTE ! Screw (3) is locked with locking
varnish. It can be released by a sharp tug using a
hexagon socket key. Apply more locking varnish to
screw (3) and tighten it.
The closing panel (6) can also be adjusted in the
direction of the arrow after releasing the screw (5). It is
important that the panel is then screwed down firmly.
Figure 9 Layout of the door hinge
18.1 CLEANING
Regular cleaning of the easy-to-clean inside of the chamber prevents deposits which over time can
detract from the appearance and the functionality of the stainless steel chamber .
The metal surfaces of the oven can be cleaned with commercially available cleaning agents for
stainless steel. It is important to ensure that no rust-forming object comes into contact with the
chamber or the stainless steel casing. Rust deposits cause infection of the stainless steel.
If any contamination causes rust stains on the surfaces of the chamber, such spots must be cleaned
off immediately and the area polished. The control panel, the plastic input modules and other plastic
components of the heated protective housing must not be cleaned using scouring cleaning agents or
those contained solvents.
19 CLOSING DOWN
Before closing down, i.e. switching off the heating, the heated gas conditioning system and if
necessary the interior of the heated protective housing should be purged with inert gas or air in order
to avoid a condensation and an eventual acidification.
20 ERROR MASSAGES
E-0 Error on self test
E-1 Power module triac faulty
E-2 Power module faulty
E-3 PT100 temperature probe faulty