Technical Data
W A R N I N G ! I M P O R T A N T !
An external fine filter must always be used at the gas inlet of the analyser. Depending
on the composition of the sample gas, it may be necessary to use a sample conditio-
ning system. Without precautions, the analyser is only suitable for measuring of non-
hazardous gases or gasmixtures in non-hazardous areas.
Version PMA 30 thermostated oxygen analyser in compact 19-inch cassette
Part No. PMA 30/D: 03A2005 = 230V 50Hz, 03A2005a = 115V 60Hz
PMA 30A/D: 03A2001 = 230V 50Hz, 03A2005a = 115V 60Hz
Measuring ranges selectable for 0-1, 0-3, 0-10, 0-30 und 0-100 Vol.% O
2
, linear
Expanded measuring ranges***
option: zero suppression for spreading of measuring range, adjusted by the standard
ranges, for example 20-21% (1%), 18-21% (3%), 90-100% (10%), max.97-100%, incl. process pressure compensation
type SDPD** Part No.: 03A9535
External measuring range indication potential free contacts for each selected range, contact rating max. 48V DC, 0,5A
Remote range control Remote range control for all 5 ranges by means of external potential free contacts
Indication
PMA 30A/D: analogue meter with a scale of 0-30 and 0-100% for each selected range,
digital meter, 4
1
/
2
digit 9 mm high LCD-indicator for 0-100% O
2
reading, selectivity 0,1vol.% O
2
PMA 30/D: digital meter 4
1
/
2
digit 18 mm high LCD for 0-100% O
2
reading, selectivity 0,1 vol.% O
2
Output signals 0-10V DC non-isolated, load > 100 KΩ, for the range of 100 vol.%, and
0-20mA* or 4-20mA* for the chosen range, non-isolated, max. load 300Ω / 900Ω
Response time for 90% FSD < 3 seconds at 60 Nl/hr air
Accuracy after calibration deviation:
analogue signal output = ±1% of span at range 3-100% / digital indicator = ±0,1 vol.% O
2
= ±2% of span at range 1%
Reproducibility deviation: analogue = < 1% of span / digital = ± 0,1 vol.% O
2
Influence of ambient temperature no influence up to 45 °C
Influence of barometric pressure the oxygen reading varies in direct proportion to changes of the barometric pressure
Influence of sample gas flow variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1 vol.% O
2
Sample gas inlet pressure 0,01 up to 1 bar g, (PMA 30 required admission pressure for competent flow rate, no pump inside)
Sample gas outlet pressure outlet of analyser must discharge freely into atmosphere;
option: process pressure compensation PD** in a range of 0,4 to 1,6 bar absolut Part No.: 03A9530
Flow rate of sample gas max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70 Nl/hr
Temperature of sample gas -10 °C up to +40 °C dry gas
O
2
-transducer temperature fixed at +50 °C
Ambient temperature -10 °C up to +45°C
Storage temperature -20 °C up to +60 °C, relative humidity 0-90% RH
Power supply internal power unit for 230VAC standard or 115VAC available (a)* +/-15%, 40-60Hz, 35,5VA
Electrical connections mains supply: 3-pole chassis plug with 2 m of cable; signals: 9 and 25-pole plugs
Materials in contact with sample gas Platinum, Glass, PTFE, PVDF, Stainless Steel 316Ti, FPM, Epoxy resin
Sample gas connections 1/8" NPT i, optional hose connector DN 4/6 available
Flow alarm thermo-conductive flow sensor downstream mounted after measuring cell
Status alarm for low sample flow, transducer temperature < 40 °C, light beam defect, power failure:
LED indication and potential free change over contact, contact rating max. 48V DC, 0,5A (2x)
Protection / electrical standard IP 20 EN 60529 / EN 61010
Housing / front colour 3U/28HP - 19-inch cassette European standard / silver
Dimension / weight height 3U, width 28HP, length 230 mm + approx. 60 mm installation space / approx. 2,5 kg
Options
03A9150 Wall mountage housing out of aluminium type 30-19 with 19“-insertion system (3U/42TE), incl. clamp box, clamp
boards and filter, with window. protection class: IP65.
* Please specify with order.
** Without SIL-certication
*** Standard with analogue and digital meter.