Technical information
Table Of Contents

46 
• use oxide-sharpened Si
3
N
4
 probe tips, e.g. Model NPS which has the following specifications 
(http://www.di).com/products2/NewProbeGuide/ContactModeProbes.html : 
PARAMETER VALUE 
spring constant  0.58 N/m 
nominal tip radius of curvature  5-40nm 
cantilever leg length 
100µm 
cantilever configuration  V-shaped 
reflective coating  Au 
shape of tip  square pyramidal 
tip half angle  35
o
• use piezo scanner with smallest distance range available, A or EV scanner, E scanner should also work (the 
difference is the engagement mechanism). 
• load sample, turn microscope and vibration table on, and then focus the laser as far out on the cantilever as 
possible to obtain the highest sensitivity 
• let the system stabilize thermally for 30 minutes (with hood on) 
• laser focusing (find maximum): the laser should never been switched off, i.e. turn the system on and off and 
all manipulations done with the laser on 
PANEL PARAMETER SETTING 
Scan size 
1 µm 
X offset  0.00 nm 
Y offset  0.00 nm 
Scan angle  0.00 deg 
Scan rate  61.00 Hz 
Number of samples  512 
Slow scan axis  enabled 
Scan Controls 
Z limit  between 55V-440V 
Integral gain  0.001 
Proportional gain  0.00 
Lookahead gain  0.00 
Feedback Controls 
Setpoint 0 V 
AFM mode  Contact Other Controls 
Input attenuation  1x 
Interleave Controls  Interleave mode  Disabled 
Channel 1  Data type  Height* 
(*setting this parameter to 
deflection is typically easier) 
Highpass filter  OFF, 3-4  
Lowpass filter  OFF, 1 
III. Imaging 
• Engage the surface. Make sure you are not false engaged (*see Section 10.10.1 of the DI AFM Manual). 
• Reduce 
Scan Size
 to ~12 nm. 
OR 
• Engage with the Scan Size set to zero and slowly increase. 
• Increase Scan Rate to 60Hz. Notice that if the Scan Rate is set much higher for atomic scale images to 
defeat some of the noise due to thermal drift. 
• Adjust Integral Gain, Setpoint, Scan Rate, and Scan Angle to obtain a good image. Initially, the Setpoint 
should be kept as low as possible initially and then increased to obtain an image. The Z-center position 
should be close to 0V. The Scan Angle is known to have a huge effect, with optimal imaging conditions if the 










