Hardware manual

Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-137
Figure 112 Select Use Dead Time Calibration.
4 Introduce the 50ppb Er solution.
Tune the sensitivity to make the count of the major isotope
166
Er in the
range of 50,000 - 1,000,000 count / 0.1 sec.