Appendix D: Remote Conguration Flow Charts
D.1 ADVANCED CONFIGURATION
Command - ALL0
Turn off all alarms while configuring unit
Command - DIN1#
Select what instrument is on TSC1
Command - TSF1#
Set the scale factor for TSC1
Command - R
Reset unit to known state
Command - SFC#
Enable or disable cross loop function
Command - SFT#
Enable or disable tandem function
Command - DIN2#
Select what instrument is on TSC2
Command - UI#
Set the input units for TSC1
Command - TSF2#
Set the scale factor for TSC2
Command - BT#
Set brake type on TSC1
Command - G#
Select max excited speed of PB
START
A
TSC not selected yet
Command - ALA#
Enable or disable air flow alarm
Command - ALE#
Enable or disable external alarm
Command - ALW#
Enable or disable water flow alarm
Command - TSC1 or TSC2
Select TSC to control
Command - AF1#
Select filter for TSC1
Command - AF2#
Select filter for TSC2
Command - ALL1
Turn on all alarms
A
DONE
WAIT 6 SECONDS