User`s manual

26
Magtrol Model DSP6001 Dynamometer ControllerChapter 3 – Installation/Configuration
SETUP
3.2.6 eDDy-CUrrent or powDer Brake DynaMoMeter SetUp
3.2.6.1 Hardware Connection
DYNAMOMETER / TSC1
AUX / TSC2
SUPPLY 2
SUPPLY 1
RS-232C
GPIB/IEEE–488
BRAKE
ACCESSORY
TORQUE–SPEED
OUTPUT
CTRL OUT
BRAKE FUSE
CAUTION: DOUBLE POLE FUSING
75VA 50/60Hz
EARTH
GROUND
FUSE (5×20mm):
120V UL/CSA 800mA 250V SB
240V IEC 315mA 250V T
(5×20mm):
UL/CSA 1.25A 250V SB
IEC 1A 250V T
MAGTROL, INC. BUFFALO, NY
GPIB
or
RS-232
PC
M-TEST
Excitation
Torque
TSC 401
Torque-Speed
Conditioner
DES Power Supply
Speed
Eddy-Current (WB)
OR Powder Brake (PB)
Dynamometer
DSP6001 Dynamometer Controller
Figure 3–20 Eddy-Current or Powder Brake Dynamometer Setup
3.2.6.2 Software Configuration
1. Turn on the DSP6001 and proceed to the dynamometer configuration menu. See Section
3.2.1 – Dynamometer Configuration Menu.
2. Select TSC1 until WB or PB is reached.
3. Select TSC2 until AUX is reached.
4. Press SHIFT. The display should appear as follows:
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
Figure 3–21 TSC1 Eddy-Current/Powder Brake Setup Menu
5. Press TORQUE UNITS button and use UP and DOWN buttons and Decrease/Increase
dial to set desired scale factor.
6.a. If using an Eddy Current Dynamometer, press SHIFT 2 times. The display should appear
as shown in Figure 3–19 Nominal Speed Setup Menu. Press the AUX SETUP button and
use UP and DOWN buttons and Decrease/Increase dial to set desired nominal speed.
Press SHIFT 4 times to complete the initial setup and return to the main menu.
6.b. If using a Powder Brake Dynamometer, press SHIFT 4 times to complete the initial setup
and return to the main menu.