Chapter 3 – Installation/Configuration
SETUP
3.2.11 eDDy-CUrrent DynaMoMeter witH powDer Brake DynaMoMeter (tanDeM SetUp)
3.2.11.1 Hardware Connection
DYNAMOMETER / TSC1
AUX / TSC2
SUPPLY 2
SUPPLY 1
RS-232C
GPIB/IEEE–488
BRAKE
ACCESSORY
TORQUE–SPEED
OUTPUT
CTRL OUT
BRAKE FUSE
CAUTION: DOUBLE POLE FUSING
75VA 50/60Hz
EARTH
GROUND
FUSE (5×20mm):
120V UL/CSA 800mA 250V SB
240V IEC 315mA 250V T
(5×20mm):
UL/CSA 1.25A 250V SB
IEC 1A 250V T
MAGTROL, INC. BUFFALO, NY
GPIB
or
RS-232
PC
M-TEST
Clutch
(EK)
DSP6001
Dynamometer
Controller
Transformer
Powder Brake
Dynamometer (PB)
Excitation
Torque
DES Power Supply
TSC 401
Torque/Speed
Conditioner
Eddy-Current Brake
Dynamometer (WB)
Excitation
Torque
TSC 401
Torque-Speed
Conditioner
DES Power Supply
Speed
Figure 3–27 Eddy-Current Dynamometer with Powder Brake Dynamometer (Tandem Setup)
3.2.11.2 Software Configuration
1. Turn on the DSP6001 and proceed to the dynamometer configuration menu. See Section
3.2.1 – Dynamometer Configuration Menu.
2. Select TSC1 until WB is reached.
3. Select TSC2 until PB is reached.
4. Press SHIFT to get to the TSC1 Eddy-Current Setup Menu as seen below.
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
Figure 3–28 TSC1 Eddy-Current Setup Menu
5. Press TORQUE UNITS button and use UP and DOWN buttons and Decrease/Increase
dial to set desired scale factor for TSC1.
6. Press SHIFT 2 times. The display should appear as shown in Figure 3–19 Nominal Speed
Setup Menu.
7. Press the AUX SETUP button and use UP and DOWN buttons and Decrease/Increase
dial to set desired nominal speed.