Magtrol Model DSP6001 Dynamometer ControllerChapter 6 – Alarm System
SETUP • OPERATION
6.1.3 aLarM priority
While in an alarm condition, a higher priority alarm will be acknowledged, while lower priority
alarms are ignored. The priority order is as follows.
Priority Alarm
Availability
Hysteresis
Dynamometer
Eddy-Current/Powder
Brake Dynamometer
1 Electrical Alarm N/A X
2 Temperature Alarm N/A X
3 External Alarm X X
4 Air Flow Alarm X N/A
5 Water Flow Alarm N/A X
6 Maximum Torque X X
7 Maximum Speed X X
8 Power X X
6.2 POWER ALARM
• Usedtoindicateanoverpowercondition
• Defaultissetat1kW
6.2.1 inStrUCtionS for power aLarM SetUp
1. Starting from main menu, press SHIFT.
2. Press DYNO SETUP button.
3. Select MAXPOWER. The display should appear as follows:
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
Figure 6–5 Max Power Menu
4. Press MAX SPEED button and use UP and DOWN buttons and Decrease/Increase
dial to set desired maximum power value for TSC1.
5. Press SHIFT.
6. Press POWER UNITS button, then MAX SPEED button and use UP and DOWN
buttons and Decrease/Increase dial to set desired maximum power value for TSC2.
7. Press SHIFT 3 times to complete Power Alarm setup and return to the main menu.