User`s manual

53
Magtrol Model DSP6001 Dynamometer Controller
Chapter 6 – Alarm System
SETUP • OPERATION
6.2.2 power aLarM aCtion
When the power exceeds that of the maximum power setting, the message -OL- will appear and
blink in the power section of the display accompanied by an audible beeping sound (as indicated
in Figure 6–6).
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
Figure 6–6 Power Alarm Display
Because torque and speed have separate setable limits, no other action will be taken. If condition
persists the temperature sensors will open causing the temperature alarm to activate.
6.2.3 to reSet power aLarM
This is a non-latching alarm. When the condition goes away, the alarm will automatically stop.
6.3 MAXIMUM SPEED ALARM
• Usedtolimitspeedofsystem(motor,dynamometer,couplings,etc.)
• Defaultissetat4000rpm
6.3.1 inStrUCtionS for MaxiMUM SpeeD aLarM SetUp
1. Starting from main menu, press SHIFT.
2. Press DYNO SETUP button.
3. Select ENCODERS. The display should appear as follows:
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
Figure 6–7 Speed Alarm Setup Menu
4. Press MAX SPEED button and use UP and DOWN buttons and Decrease/Increase
dial to set desired maximum speed for TSC1.
5. Press SHIFT.
6. Press POWER UNITS button, then MAX SPEED button and use UP and DOWN
buttons and Decrease/Increase dial to set desired maximum speed for TSC2.
7. Press SHIFT 3 times to complete Maximum Speed Alarm setup and return to the main
menu.