User`s manual

59
Magtrol Model DSP6001 Dynamometer Controller
Chapter 6 – Alarm System
SETUP • OPERATION
6.8 TEMPERATURE ALARM
• Toalertuserwhendynamometergetstoohotandthermalswitchopens
• OnlyavailableforusewithWBorPBdynamometers
• Default-alwaysactive
6.8.1 inStrUCtionS for teMperatUre aLarM SetUp
No setup needed.
6.8.2 teMperatUre aLarM aCtion
When the dynamometer in use becomes too hot, the display will flash “TEMPERATURE ALARM
TSCX (as indicated in Figure 6–19) and will be accompanied by an audible beeping sound.
The alarm relay will open and the excitation current will immediately decrease to 10%. After
approximately 3 seconds, the current wll drop to 0.
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
∗∗ ∗∗
Figure 6–19 Temperature Alarm Message Display
6.8.3 to reSet teMperatUre aLarM
Press any front panel button other than SHIFT. The alarm condition must be cleared before the unit
will return to normal operation.