Magtrol Model DSP6001 Dynamometer ControllerChapter 6 – Alarm System
SETUP • OPERATION
6.9 ELECTRICAL ALARM
• UsedtoprotecttheDESsupply
• Monitorselectricalinput(mains)andcircuitryoftheDES
• OnlyavailableforusewithWBorPBdynamometers
• Default-alwaysactive
6.9.1 inStrUCtionS for eLeCtriCaL aLarM SetUp
No setup needed.
6.9.2 eLeCtriCaL aLarM aCtion
When there is an electrical overload, the display will flash “ELECTRICAL ALARM TSCX” (as
indicated in Figure 6–20) and will be accompanied by an audible beeping sound. The alarm relay
will open and the excitation current will immediately decrease to zero.
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
∗∗ ∗∗
Figure 6–20 Electrical Alarm Message Display
6.9.3 to reSet eLeCtriCaL aLarM
Press any front panel button other than SHIFT. The alarm condition must be cleared before the unit
will return to normal operation.