User`s manual

79
Magtrol Model DSP6001 Dynamometer Controller
Chapter 9 – Calibration
MAINTENANCE
15. N/C
14. CURRENT SET POINT (ANALOG 0V)
13. +5.0 VDC COM
12. SUPPLY +24 VDC
11. STAND-BY CHANNEL 1
10. TEMPERATURE ALARM CHANNEL 1
9. SHIELD (EARTH)
8. W FLOW_1
7. CURRENT SET POINT (SIGNAL)
6. +5.0 VDC COM
5. SUPPLY +24 VDC
4. PRIMARY SUPPLY CONTR. CHANNEL 1
3. N/C
2. ELECTRICAL ALARM CHANNEL 1
1. SHIELD (EARTH)
14 3 25
912 11 10
68 7
14 1315
Figure 9–5 Supply 1 Connector
3. Using the Decrease/Increase dial, adjust the value so that the meter reading is 0.000 volts
or within 00.20 mV.
4. Press SHIFT once.
5. Connect meter to Pin 14 (negative) and Pin 7 (positive) of SUPPLY 2 connector.
3. CLUTCH
8. W FLOW_2
11. STAND-BY CHANNEL 2
1. SHIELD (EARTH)
2. ELECTRICAL ALARM CHANNEL 2
4. PRIMARY SUPPLY CONTR. CHANNEL 2
5. SUPPLY +24 VDC
6. +5.0 VDC COM
7. CURRENT SET POINT (SIGNAL)
9. SHIELD (EARTH)
10. TEMPERATURE ALARM CHANNEL 2
15. N/C
13. +5.0 VDC COM
14. CURRENT SET POINT (ANALOG 0V)
12. SUPPLY +24 VDC
14 3 25
912 11 10
68 7
14 1315
Figure 9–6 Supply 2 Connector
6. Using the Decrease/Increase dial, adjust the value so that the meter reading is 0.000 volts
or within 00.20 mV.
7. Press SHIFT once. The display should appear as follows:
POWER
BRAKE STATUS SET POINT SET POINT P I D
TORQUE SPEED STATUS
Figure 9–7 Calibration Accessory Torque DAC Display Menu