User`s manual

81
Magtrol Model DSP6001 Dynamometer Controller
Chapter 9 – Calibration
MAINTENANCE
9.3.6 DeCiMaL point CHeCk
1. Connect a 0.5 VDC source to pins 14 (TORQUE SIGNAL) and 13 (TORQUE COMMON)
of the 14-pin TSC1 connector. For reference, see Figure 9–2 TSC1 Input Connector.
2. Verify that the torque display reads 500.0.
3. Connect pin 9 (D.P. A) to pin 8 (+5.0 VDC COM).
4. Verify that the torque display reads 50.0.
5. Disconnect pin 9 from pin 8.
6. Connect pin 12 (D.P. B) to pin 8 (+5.0 VDC COM).
7. Verify that the torque display reads 5.000.
8. If the torque display does not show correct readings, call Magtrol Customer Service.
Note: There are no adjustments for decimal point location.