Specifications

24
7 - T7 - T
7 - T7 - T
7 - T
roubleshootingroubleshooting
roubleshootingroubleshooting
roubleshooting
If you require additional assistance, please contact
Magtrol Customer Service at 1-800-828-7844 or 1-716-668-5555
PROBLEM REASON SOLUTION
Display indicates I/O ERROR
Command does not match the
unit's pro
g
rammed set of
instructions.
Use correct command and format.
Speed command sent, but
Controller does not respond.
Communication occured but the
Controller is not loadin
g
the motor.
Adjust PID values.
Mechanical power reads much
hi
g
her or lower than expected.
Torque units are incorrect.
Set torque input units to match the
specifications on dynamometer nameplate.
No GPIB communication Setup error and/or hardware fault.
Check:
GPIB address of Controller
GPIB cable: should be functionin
g
and
attached to Controller and computer
interface card.
No RS-232 communication Setup error and/or hardware fault.
Check:
Baud rate of Controller
Pinout of serial cable
Cable attachment to Controller and serial
interface port of computer.
Dynamometer shaft does not
turn smoothly when BRAKE is
OFF.
Salient poles were set up on the
rotor by havin
g
brake current
applied with no shaft rotation.
Start the motor and brin
g
up to speed.
Press BRAKE button ON. Adjust output
current up to a value at least 25% of the
maximum torque ratin
g
of the dynamometer
in use (if possible). Reduce output current
to 0.