User`s manual

44
Magtrol WB/PB DynamometersChapter 2 – Installation / Mounting
2.1.3 WB/PB 65 SerieS
WB/PB 65 Series Dynamometers are fitted with a mechanical stator assembly locking device
protecting the load cell during transport. After installing the dynamometer on the test bench, proceed
as follows and refer to figure 2–3 to unlock the system:
1. Loosen the two nuts
.
2. Unscrew the two screws
with three turns.
3. Tighten the two nuts
.
Note: The screws and nuts for overload protection and are adjusted
at the factory (refer to Chapter 6 Calibration for possible
adjustments).
Figure 2–3 WB/PB 65 Load Cell Locking System