User`s manual

45
Magtrol WB/PB Dynamometers
Chapter 2 – Installation / Mounting
2.1.4 WB/PB 115 SerieS
WB/PB 115 Series Dynamometers are fitted with a mechanical stator assembly locking device
protecting the load cell during transport. After installing the dynamometer on the test bench, proceed
as follows and refer to figure 2–4 to unlock the system:
1. Loosen the two nuts
.
2.
Unscrew the two screws with three turns.
3.
Tighten the two nuts .
Note: The screws and nuts for overload protection and are adjusted
in the factory (refer to Chapter 6 Calibration for possible
adjustments).
Figure 2–4 WB/PB 115 Load Cell Locking System