6. Calibration
Every new Magtrol WB/PB Series Dynamometer is factory calibrated. Static calibration of the zero
and torque is recommended from time to time, especially after having maintenance performed.
Note: Do not forget to unlock the dynamometer following transport (after
it returns from overhaul) in accordance with Section 2.1 Locking
the Dynamometer for Transport.
6.1 BASIC CALIBRATION INFORMATION
Prior to calibration, the motor must be separated from the dynamometer so that the dynamometer
shaft can rotate freely.
6.1.1 c
AliBrAtion ArmS
Two calibration arms are delivered with each Magtrol WB/PB Series Dynamometer. These arms
are inserted into the threaded holes found on each side of the stator. See Figure 6–1 Calibration
Arm Mounting.
Each calibration arm has two grooves: the outer groove allows for calibration in m, the inner in mkp.
Note: Mkp is an obsolete unit of torque measurement. Magtrol recommends
calibration in N·m (by using the outer groove) and calibration weight
in kg. Refer to Section 6.1.2 – Calibration Weight.
4/2 15%0/7 %2 30%%$
#OL DWA TER
INL ET
4HREADEDHOLEFORATTACHING
THECALIBRATIONARM
2EMOVABLECALIBRATIONARMS
#ALIBRATIONWEIGHT
INKG
;.qM=
;MKP=
Figure 6–1 Calibration Arm Mounting
Caution: after Calibration, do not forget to remove the Calibration arms
from the dynamometer.