Instruction manual

Instruction Sheet
for the PASCO
Model ES-9053A
012-04221C
11/94
$1.00
RESISTOR -CAPACITOR NETWORK
Introduction
The PASCO Model ES-9053A Resistor-Capacitor Net-
work is designed for the study of RC time constants and
for investigating capacitors connected in series and paral-
lel. The RC time constants range from 25 to 200 seconds,
so charging and decay curves can easily be investigated
using an electrometer, such as the PASCO Model
ES-9054B Student Electrometer, and a stopwatch.
Two resistors and three capacitors are wired into a
switching network, which allows selecting three resis-
tance values (50 ΜΩ, 100 M, and 200 M) in series
with two selectable capacitance values (0.47 µF and .94
µF). A third capacitor (0.22 µF) can be selected in series
or parallel with the other two capacitances.
The switches and external connections allow the follow-
ing combinations:
Six combinations of resistors and capacitors in series.
Two combinations of capacitors in series.
Two combinations of capacitors in parallel.
Three slide switches determine the component selection
and circuit configuration (charging, discharging, or open cir-
cuit). The positions of these slide switches on the front panel
diagram indicate the exact component values being used.
1.0 x 10 ohms
8
.47 x 10 farads
-6
(B)
(A)
Figure 1: Switch Positions & Equivalent Circuit
Figure 1 (A) shows the slide switches in arbitrary posi-
tions, while Figure 1 (B) reflects the corresponding
equivalent circuit diagram.
The three position slide switch on the left (S3) has a
very useful OPEN position. It is often convenient to
stop the charging or discharging process when making
a measurement, and the OPEN position permits the
circuit to be disconnected.
OPEN
COMPONENTS: ±5%
MAX. 50 VOLTS
1.0 X 10
8
2.0 X 10
8
.50 X 10
8
.22 X 10 F
-6
.47 X 10 F
-6
.94 X 10 F
-6
EXT
INPUT
GND
IN
GND
OUT
ES-9053A
RESISTOR - CAPACITOR
NETWORK
OPEN
COMPONENTS: ±5%
MAX. 50 VOLTS
1.0 X 10
8
2.0 X 10
8
.50 X 10
8
.22 X 10 F
-6
.47 X 10 F
-6
.94 X 10 F
-6
IN
OUT
GNDGND
EXT
INPUT
© 1990 PASCO scientific