User's Manual

www.malags.com
26
7 Power supply, Optical fibres and
Communication cables
_________________________________________________
7.1 Power supply
Each electronic component in the RAMAC/GPR system is
powered by specially designed battery packs. These packs are
interchangeable between system components.
Rechargeable Ni-Cd or Pb-acid battery packs are available.
The capacity of the Ni-Cd packs is approximately 5.7 Ah. The
Pb-acid battery packs have an approximate 8 Ah capacity.
The operating time of the system is dependent on the charge
cycle history of the batteries. Normally, maximum operating
time is not reached until the batteries have been fully charged
and discharged 3-5 times. Optimum performance is achieved
through fully discharging and recharging the battery packs.
Note that batteries lose efficiency in cold temperatures. So
insulating the electronics and battery packs in cold climates
will prolong the battery life as the electronic units generate
internal heat during operation.
The batteries are mounted by simply placing the battery
compartment with the lid on the rear short side under the
corresponding groove in the electronic units. The front end or
the battery pack has a locking tab on the plastic housing.
Gently press on the front end of the battery until the lid is
released from the groove before removal.
A special empty battery pack for the CUII and unshielded
antennas is also available. It can hold six standard D-size-
cells. These can easily be replaced and are suitable when