Specifications
23G-Series RO System
3024270 Rev G 97 14Oct13
• Restore the offset settings and wiring when the new transducer is
installed.
• The new transducer must be calibrated (see Section 8.6).
• Remove transducer and clean port.
If this remedy does not work, contact your equipment supplier.
7.1.6 Permeate Flow (Stuck at Zero)
• Check flow turbine to ensure it spins freely.
• First check connections (Figure 4.6), power down for 2 minutes,
recall Factory Settings and check the values in the Calibration
Menu (06).
• Switch sensor cables, or insert the permeate cable into the
concentrate turbine (the machine will run normally with a
concentrate flow reading of zero).
• Unplug the cable from its turbine, and set the flow offset in the
Calibration Menu to 99 (forces the flow reading to 0.1), then set
the permeate low flow alarm to 0.0.
• Return to the proper settings after the problem is corrected.
• Remove turbine and blow through it to be sure it spins freely.
If this remedy does not work, contact your equipment supplier.