Specifications

23G-Series RO System
3024270 Rev G 117 14Oct13
4. The “Conc Flow rate Offset” should read 00. If it does not read
00:
Enter the passcode in the following sequence to adjust the
reading:
a. HAND/AUTO
b. CIP
c. ALARM BYPASS
Press the ARROW keys to set the "Conc Flow rate Offset"
to 00.
Press ENTER to save the new value.
The Concentrate Flow Rate Offset is now calibrated.
Calibrating Permeate Flow Rate Offset
NOTE:
It is not necessary to enter the passcode for additional adjustments in the
Calibration Menu.
1. Press ENTER until the “Perm Flow rate Offset” is displayed.
2. The “Perm Flow Rate Offset” should read 00. If it does not read
00:
Press the ARROW keys to set the “Perm Flow rate Offset”
to 00.
Press ENTER to save the new value.
The Permeate Flow Rate Offset is now calibrated.
8.7.4 Verifying the Concentrate and Permeate Flow Rate Slopes
Verifying the Concentrate Flow Rate Slope
NOTE:
It is not necessary to enter the passcode for additional adjustments in the
Calibration Menu.
1. Press ENTER until the “Conc Flow Rate Slope” is displayed.