Specifications

23G-Series RO System
3024270 Rev G 118 14Oct13
2. The “Conc Flow Rate Slope” should read 37. If it does not read
37:
Press the ARROW keys to set the “perm Flow Rate Slope”
to 37.
Press ENTER to save the new value.
The above steps verify the factory default sets for the concentrate
and permeate flow rate slopes. The following procedure will
calibrate the slopes.
Verifying the Permeate Flow Rate Slope
NOTE:
It is not necessary to enter the passcode for additional adjustments in the
Calibration Menu.
1. Press ENTER until the “Permeate Flow Rate Slope” is
displayed.
2. The “Perm Flow Rate Slope” should read 37. If it does not read
37:
a. Press the ARROW keys to set the “Perm Flow Rate Slope
to 37.
b. Press ENTER to save the new value.
c. The above steps verify the factory default sets for the
concentrate and permeate flow rate slopes.
8.7.5 Calibrating the Concentrate and Permeate Flow Rate Slopes
Calibrating the Concentrate Flow Rate Slope
1. Press ESC twice to return to the ECM-100 default menu.
2. Press the ARROW keys to scroll to the Calibration Menu (06).
3. Press ENTER to select the Calibration Menu.
4. Press ENTER to scroll to the “Conc Flow Rate Slope.”