Specifications
23G-Series RO System
3024270 Rev G 34 14Oct13
4.2.6 ECM-100 Passcode
Your machine is equipped with an ECM-100 controller/monitor to
perform the machines electrical operations. Selected functions of
the ECM-100 require entry of a passcode. This passcode access
allows three functions; it allows a user to make adjustments to
setpoints, view, and download historical data. The passcode is
intended to provide access adjustment capability to authorized
personnel. The passcode for your ECM-100 is entered by pressing
the function keys on the ECM-100 keypad, in the following
sequence:
1. HAND/AUTO
2. CIP
3. ALARM BYPASS
Contact your equipment supplier for assistance.
4.2.7 Factory Settings Storage
1. The “factory settings” is a copy of all menu data stored in a
separate reserved storage area.
2. This data can be updated by following the procedure described
as “Store Factory Settings.”
3. Mar Cor Purification factory personnel saved a copy of menu
settings before the machine was shipped.
4. At installation, after timers and calibrations have been tailored
for your particular application, and/or after recalibration of pH
probe, the settings need to be copied to the factory settings
area for permanent storage.
5. In the event that the settings have been corrupted, the menu
settings can be recalled.
6. To store Factory settings:
a. Go to the Factory Settings Menu (menu 11).
b. Push ENTER switch. The screen will display two lines of
factory settings.
c. Enter storage code in the following sequence:
HAND/AUTO
CIP
ON
ENTER