Specifications
23G-Series RO System
3024270 Rev G 37 14Oct13
08 Print Current Values
The Print Current Values Menu is used to facilitate a download of
current operational data.
Press ENTER to print/send the data that is currently displayed on
the ECM-100 screen to a printer or computer.
09 Metric/U.S. Select
The Metric/U.S. Select Menu option is used to specify the units of
measure displayed on the ECM-100.
Press ENTER, then use the ARROW keys to toggle between
choice of units. Press ENTER to select.
10 Perm Hi Cond Alarm
The Perm Hi cond Alarm Menu is used to specify/modify the
machine’s response to a high permeate conductivity condition.
Enter the passcode to select one of the three alarm responses. Use
the ARROW keys to toggle through the responses. Press ENTER
to select.
11 Factory Settings
The Factory Settings Menu is used to store and recall menu
settings.
Enter the passcode to recall default factory settings. See Section
4.4 for saving and recalling factory settings.
12 Auto-On Tank Full
The Auto-On Tank Full Menu is used to control the machines
response to a tank full condition.
Press ENTER, then use the ARROW keys to toggle between
shutdown (default) and divert (used only if permeate divert option is
installed). Press ENTER to select or ESC to exit.
NOTE:
If “Divert” is selected, the AUTO-ON function will come on even if the tank is full,
and Perm Divert will come on anytime the tank is full (even in HAND mode).
NOTE:
Appropriate plumbing must be installed to divert permeate and to prevent
overflowing the tank. The machine should be left “ON” and in AUTO. After the
tank fills up normally, the AUTO-ON will periodically start-up and run for a short
time to flush the machine. During this time the permeate is in DIVERT mode.