Specifications
23G-Series RO System
3024270 Rev G 62 14Oct13
3. Hook up the CIP feed hose to the quick connect fitting on the
pump. Hook the other end of the hose to the quick connect
fitting on the CIP tank.
4. Place the CIP skid near the 23G so all CIP lines will reach
CIP ports on the machine.
5. Connect the feed permeate and concentrate lines to the CIP
tank quick connect on CIP pump and then to the CIP port on
the machine. Open the CIP valves on the machine.
5.8.2 Cleaning/Sanitizing with 23G CIP
1. Start the 23G and allow the permeate to fill the 23G skid
tank to the desired level. Turn machine off by pushing the
Off button on the ECM-100 controllers.
2. Add and mix desired amount of cleaner or sanitizer to the
CIP tank.
3. Mix chemical cleaning solution in a suitable (clean) tank with
purified water. RO permeate is best. Use the following to
determine cleaning solution volume:
Cleaning Solution Volume (Permeate Flow + Concentrate Flow) x3
Permeate and Concentrate are specified in gpm (Lpm)
Cleaning solution volume will be in gallons (L)
4. Remove the feed CIP line from the quick connect fitting on
the side of the CIP pump. Connect it to the quick connect
fitting on the CIP feed port on the side of the 23G.
5. Connect the concentrate CIP line to the quick connect port
on the CIP skid. Attach the other end to the quick connect
concentrate CIP port on the 23G.
6. Close the feedwater valve which provides water from
pretreatment to the machine.
7. Close the permeate and concentrate outlet valves on the
back side of the front panel.
8. Then connect the long cord from the starter box on the CIP
skid to the pigtail bell fitting on the back side of the controller
box.
9. Connect 115-volt power to the Medical cord, located on the
starter box on the CIP skid.