Specifications
23G-Series RO System
3024270 Rev G 73 14Oct13
all the preservatives have been removed from the
membrane elements and prevent irreversible
damage. If the 23G RO machine is going to be
disinfected, rinse the system to drain under normal
pressures for a minimum of one (1) hour before
exposing membranes to Minncare
®
.
Flush the system to drain under normal pressures for
a minimum of six (6) hours before exposing
membranes to formaldehyde.
Machine is ready for normal operation, however do
not put into service until all applicable water tests
have been completed.
In order to ensure that the RO is performing at or
above AAMI standards, it is necessary to perform an
AAMI Water Quality Analysis Test. For new units, the
RO must be operated (rinsed to drain) for at least two
hours before taking a sample, to ensure that the
membranes are free of preservatives or other
contaminants which would invalidate the test.
NOTE:
Follow the procedures recommended by your water quality testing
laboratory.
5.15 Flow Control Adjustments
NOTE:
Prior to any adjustments, make sure that temperature is constant,
AUTOFLUSH is off, and concentrate valve is closed.
Refer to Figure 5.11, Flow Control Manifold.