CAUTION:
The flow control orifice valves are sealed to prevent tampering, and should
not be adjusted by unauthorized personnel. Damage to the machine may
occur and physical injury is possible.
The flow control orifices are factory set; according to the design
specifications of the machine.
Adjustment of the orifices in the field are permitted only with the
approval of the equipment supplier.
Permeate flow rates which exceed design specifications will shorten
the life of the membranes, degrade the quality of the permeate output,
and may cause injury to dialysis patients.
CAUTION:
The large orifice nuts at the base MUST NEVER be turned.
The large nuts must remain in the position set at the factory; with the
4 dimples from both nuts lined up with each other in a straight line
(longitudinal with the block).
Tightening these nuts (even by a small amount) will cause damage to
the O-ring located under the nuts.
Over-tightening the locking nuts will damage the threads and may
cause leaking the next time that the orifice is adjusted.
CAUTION:
Improper adjustment of the flow control orifices can cause serious damage
to the machine, and/or personal injury.