Specifications
23G-Series RO System
3024270 Rev G 74 14Oct13
CAUTION:
The flow control orifice valves are sealed to prevent tampering, and should
not be adjusted by unauthorized personnel. Damage to the machine may
occur and physical injury is possible.
• The flow control orifices are factory set; according to the design
specifications of the machine.
• Adjustment of the orifices in the field are permitted only with the
approval of the equipment supplier.
• Permeate flow rates which exceed design specifications will shorten
the life of the membranes, degrade the quality of the permeate output,
and may cause injury to dialysis patients.
CAUTION:
The large orifice nuts at the base MUST NEVER be turned.
• The large nuts must remain in the position set at the factory; with the
4 dimples from both nuts lined up with each other in a straight line
(longitudinal with the block).
• Tightening these nuts (even by a small amount) will cause damage to
the O-ring located under the nuts.
• Over-tightening the locking nuts will damage the threads and may
cause leaking the next time that the orifice is adjusted.
CAUTION:
Improper adjustment of the flow control orifices can cause serious damage
to the machine, and/or personal injury.