Specifications

23G-Series RO System
3024270 Rev G 81 14Oct13
6.3 Printer/Computer Interface
The ECM-100 is designed to allow data downloads to a computer or a
printer. The basic process of downloading the information to a printer or
computer is the same. Data can be downloaded as individual lines of
information (Current display), or large groups of information (Historical
log). If a computer or printer is left connected to the ECM-100 for long
periods of time, the ECM-100 will display and download (echo) each
successive set of operational data as it is stored in the historical log
table(s). The user can define the frequency of operational data storage by
accessing the Timer Presets Log Period (Menu 02) on the ECM-100.
6.3.1 Printer Interface
The ECM-100 includes a 9-pin female serial communication port
which will allow you to interface with a printer. Two print functions
are provided: Current display and History Log. See Section 6.6.6
for Print Functions.
6.3.2 Computer Interface
The serial cable on the ECM-100 is wired at the factory to connect
directly to a personal computer, requiring only a standard 9-pin
serial extension cable of appropriate length and software.
The RS-232 communication configuration is:
Baud Rate 1200, 2400, 4800, or 9600, (user must specify)
Data Length 8 bits
Parity None
Handshaking Xon/Xoff
6.3.3 Computer Connection
Software Requirements
Requires serial communications software on your computer, such
as:
HyperTerminal® (included with Windows 95, 98, NT, ME, 2000, or
XP)
Microsoft Terminal® (available separately for Windows 3.1)
Crosstalk® and Procomm® (versions available separately for
almost any computer)
- or any of a number of similar commercial products.