Specifications

23G-Series RO System
3024270 Rev G 83 14Oct13
Table 6.10 Two-Digit HyperTerminal Alarm Codes
Number on Left
Number on Right
0 No Alarm
0 No Alarm
1 Metric
1 Permeate Conductivity High
2 Permeate Divert
2 High Temperature
3 Metric and Divert
3 High pH
4 HAND mode
4 Low pH
5 Metric and HAND mode
5 Low Post Filter Pressure
6 Permeate DIVERT and HAND mode
6 High Primary Pressure
7 Metric and Divert and HAND mode
7 Low Primary Pressure
8 CIP
8 High Permeate Pressure
9 Rejection Low
A Prefilter Delta P
B Membrane Delta P
C Permeate Flow Low
D Tank Full
E Pretreatment Lockout