Specifications
23G-Series RO System
3024270 Rev G 84 14Oct13
Table 6.11 Example ECM-100 History Log
DATE
TIME
PRESSURES
FLOWS
pH
TEMP
REJ
Cp
ALARM
PRE
POS
PRI
FIN
Pp
Qc
Qp
12/10/06
15:41
41
40
260
192
56
02.2
06.8
7.6
77
96.8
04
00
12/10/06
14:41
41
39
258
192
53
02.2
06.8
7.6
75
97.4
04
00
12/10/06
13:41
40
39
258
190
54
02.2
06.7
7.5
77
97.1
04
00
12/10/06
12:41
37
35
256
188
50
02.2
06.1
7.5
71
96.2
05
00
12/10/06
11:41
41
39
260
190
56
02.2
06.7
7.5
77
96.9
04
00
12/10/06
10:41
38
37
258
190
51
02.3
06.2
7.4
71
95.4
05
00
12/08/06
20:26
42
40
260
192
58
02.1
07.1
7.5
73
96.5
04
00
12/08/06
18:26
41
40
260
192
56
02.2
06.8
7.4
78
97.3
04
40
12/08/06
15:23
05
03
132
098
39
01.5
02.8
N/A
77
97.4
04
05
12/08/06
09:44
41
39
260
190
56
02.2
06.8
N/A
78
97.5
04
00
12/08/06
07:44
40
38
258
190
55
02.2
07.1
N/A
77
97.2
04
00
12/08/06
05:44
41
39
258
190
55
02.3
06.7
N/A
77
97.6
04
00
12/08/06
03:44
42
40
260
192
56
02.3
06.5
N/A
75
96.8
04
00
12/08/06
01:41
41
39
258
190
53
02.3
06.6
N/A
75
97.3
04
00
12/07/06
23:41
41
39
258
190
54
02.3
06.6
N/A
75
96.6
04
00
12/07/06
21:41
41
39
260
190
54
02.2
06.7
N/A
75
96.7
04
00
12/07/06
19:41
41
39
260
192
55
02.2
06.8
N/A
77
96.4
05
00
12/07/06
17:12
40
38
258
190
83
02.3
01.4
N/A
73
95.8
06
08
12/07/06
15:12
42
40
258
190
57
02.4
07.1
N/A
80
96.3
04
00
12/07/06
13:12
41
39
258
190
55
02.2
07.1
N/A
78
96.4
04
00
12/07/06
11:12
42
40
258
190
56
02.3
06.8
N/A
78
97.1
04
00
12/07/06
09:12
41
39
258
190
54
02.3
06.8
N/A
77
97.4
04
00
A printout of the ECM-100 History Log provides valuable information about
the 23G. From the example above (Table 6.11):
• Permeate Pressure High Alarm at 17:12 on 12/07/06
• Inlet Pressure Low Alarm at 15:23 on 12/08/06
• The operator ran the machine in the HAND mode for a short time
on 12/08/06, and enabled the pH probe.
• On 12/10/06, before 1:40 am, the operator changed the Log Period
Timer Setting from 2 hours to 1 hour.