1.3 Alarm types
1.3.1 Notification
Next time the WRO goes to Standby the buzzer sounds, flashes,
and the message displays.
1.3.2 Info
The WRO continues in the current mode. The message text will be dis-
played.
1.3.3 Alarm
The WRO continues in the current mode. The buzzer sounds,
flashes and the message text displays.
1.3.4 Stop
The WRO stops. The buzzer sounds, flashes and the message
text displays.
1.4 Conductivity Monitoring
The conductivity monitoring system in the WRO 300 has three limits,
allowing for individual adaptation of alarm settings to local preferences
and regulations. The three limits are described in the sections below.
All three limits are set at the factory.
1.4.1 Conductivity Notification (set at 20µS/cm)
Indicates that the product water conductivity is above normal reading.
It will be issued when the WRO 300 enters Standby mode if the notifi-
cation limit has been exceeded for more than 30 consecutive seconds
during the previous run period. It may be caused by several reasons
Buzzer Sound Description
Notification
Buzzer Sound Description
Alarm
Buzzer Sound Description
Stop
NOTE
If the notification, alarm and stop limits are set to the same value the
priority of the action are in the order of stop, alarm and notification.