Operation
15
2 Operation
2.1 Start
Press until light goes on.
2.2 Stop
Press until light goes out and the WRO 300 goes into standby
mode.
WARNING
This device does not remove chlorine and chloramines. If
these substances are present in the feed water, carbon filtra-
tion is required to remove these substances. Severe patient
injury may otherwise occur. A test for total chlorine of the pre-
treated water must be performed prior to initiating dialysis
treatment. The level of total chlorine must be below 0.1 mg/l
(ppm).
NOTE
If power failure occurs during operation mode the procedure contin-
ues in the same phase as when the power failure occurred when the
WRO 300 starts up again.
RUN
X
μS
/
cm
STANDBY