Operator`s manual
P/N 3027435 Rev. C
Regular Maintenance
18
WRO 300
3.1.2 Test Kit Example
• Minncare Residual Test Strips, MCP order number 185-40-004
(preferred).
• STERICHEK
®
Residual Peroxide, MCP order number WT811905
(alternate).
3.1.3 Chemical Disinfection Procedure
Chemical disinfection of the WRO, including product water loop.
Step 1 The WRO must be in Standby mode.
Step 2 Disconnect the WRO 300 from the dialysis machine with the
quick connector on the product water loop — the connector
can be left in place. Comply with the caution message below.
CAUTION
A test for residual disinfectant after rinse must be performed
before the initiation of the next dialysis session. The residual
concentration of the disinfectant in the fluid must be below lev-
els specified by the clinic or by national standards. It is essen-
tial to use an appropriate test method, either with proven
sensitivity for the disinfectant or recommended by the manu-
facturer of the disinfectant.
CAUTION
The chemical container must be located below the chemical
intake.
CAUTION
When requested by the WRO unit, remove the disinfection wand
from the chemical intake. Failure to remove the disinfection
wand can cause siphoning of the chemical solution from the bot-
tle resulting in the tank overflowing chemical solution through the
back of the machine.
NOTE
If power failure occurs during chemical disinfection 404 RINSE
REQUIRED is displayed when the WRO 300 restarts (also 401
REMOVE WAND is displayed during the intake phase). Press until
light goes on, to initiate rinse.
CAUTION
If the WRO 300 is not disconnected from the dialysis machine
during the chemical disinfection procedure, the test for residu-
als after the chemical disinfection procedure must be per-
formed on the dialysis machine according to the Operator’s
Manual of this device.